Thread-Based Analytical Device for Determination of Phosphorus Content in Fertilizers

Authors

  • Kamonthip Sereenonchai Department of Chemistry, Faculty of Science and Technology, Thammasat University, Pathumthani 12120, Thailand
  • Chawarit Inla Department of Chemistry, Faculty of Science and Technology, Thammasat University, Pathumthani 12120, Thailand
  • Siriwit Buajarern Department of Chemistry, Faculty of Science and Technology, Thammasat University, Pathumthani 12120, Thailand

DOI:

https://doi.org/10.48048/tis.2022.5616

Keywords:

Thread-based device, Phosphorus, Colorimetry, Molybdenum blue, Cotton

Abstract

A simple microfluidic thread-based analytical device (mTAD) for determination of phosphorus content in fertilizers based on the formation of phosphomolybdenum blue (PMB) complex was proposed. The proposed mTAD was simply fabricated by stitching the 100 % cotton thread to the polyethylene (PE) mesh canvas. A color developing reagent, which are the mixture of sulfuric acid, ammonium molybdate, potassium antimony tartrate, and ascorbic acid, was firstly loaded into the middle of the thread. A sample solution was subsequently introduced into the same position to form PMB. A digital camera and image processing program were employed for evaluating color intensity of PMB. Variable parameters affecting the mTAD efficiencies were optimized i.e., solution volumes, chemical concentrations, and reaction time.  The applicability of the proposed mTAD method was evaluated by analyzing fertilizer samples. The accuracy as expressed by percentages of recovery ranged between 83 and 96 %. The method was highly reproducible with the relative standard deviation (RSD) of 0.84 % (1 mgP/L), 0.92 % (3 mgP/L) and 1.04 % (10 mgP/L) for 10 replicates. The main features of the system are the used of commonly available materials, simple and rapid fabrication, small scale, reduced chemical waste, less sample consumption, and low-cost.

HIGHLIGHTS

  • Simple materials such as 100 % cotton thread and PE mesh canvas can be used for fabricating µTAD
  • The proposed µTAD method was fast analysis and good sensitivity (as low as nmol P level within 6 minutes) by using small amount of reagent and sample solutions
  • The method was successfully determining the phosphorus content in fertilizers


GRAPHICAL ABSTRACT

Downloads

Download data is not yet available.

References

S Somnam, K Grudpan and J Jakmunee. Stopped-flow injection method for determination of phosphate in soils and fertilisers. Maejo Int. J. Sci. Tech. 2008; 2, 17281.

W Tiyapongpattana, P Pongsakul, J Shiowatana and D Nacapricha. Sequential extraction of phosphorus in soil and sediment using a continuous-flow system. Talanta 2004; 62, 765-71.

EA Nagul, ID McKelvie, P Worsfold and SD Kolev. The molybdenum blue reaction for the determination of orthophosphate revisited: Opening the black box. Anal. Chim. Acta 2015; 890, 60-82.

P Worsfold, I McKelvie and P Monbet. Determination of phosphorus in natural waters: A historical review. Anal. Chim. Acta 2016; 918, 8-20.

FTSM Ferreira, LS Mesquita, RBR Mesquitaand and AOSS Rangel. Improved sequential injection method for phosphate quantification within a wide dynamic range with in-line pre-concentration to monitor soil leachates. Talanta Open 2020; 2, 100015.

NK Ibnul and CP Tripp. A solventless method for detecting trace level phosphate and arsenate in water using a transparent membrane and visible spectroscopy. Talanta 2021; 225, 122023.

S Pradhan and MR Pokhrel. Spectrophotometric determination of phosphate in sugarcane juice, fertilizer, detergent and water samples by molybdenum blue method. Sci. World 2013; 11, 58-62.

JM Estela and V Cerdà. Flow analysis techniques for phosphorus: An overview. Talanta 2005; 66, 307-31.

BM Jayawardane, ID McKelvie and SD Kolev. A paper-based device for measurement of reactive phosphate in water. Talanta 2012; 100, 454-60.

B Waghwani, S Balpande and J Kalambe. Development of microfluidic paper basedanalytical device for detection of phosphate in water. Int. J. Innovat. Tech. Explor. Eng. 2019; 8, 592-5.

JM Racicot, TL Mako, A Olivelli and M Levine. A paper-based device for ultrasensitive, colorimetric phosphate detection in seawater. Sensors 2020; 20, 2766.

H Heidari-Bafroui, B Ribeiro, A Charbaji, C Anagnostopoulos and M Faghri. Portable infrared lightbox for improving the detection limits of paper-based phosphate devices. Measurement 2021; 173, 108607.

G Sriam, MP Bhat, P Patil, UT Uthappa, HY Jung, TAltalhi, T Kuimeria, TM Aminabhavi, R Krishna, P Madhuprasad and MD Kurkuri. Paper-based microfluidic analytical devices for colorimetric detection of toxic ions: A review. TrAC Trends Anal. Chem. 2017; 93, 212-27.

CT Kung, CY Hou, YN Wang and LM Fu. Microfluidic paper-based analytical devices for environmental analysis of soil, air, ecology and river water. Sensor. Actuator. B Chem. 2019; 301, 126855.

X Li, J Tian and W Shen. Thread as a versatile material for low-cost microfluidic diagnostics. ACS Appl. Mater. Interfac. 2010; 2, 1-6.

M Reches, KA Mirica, R Dasgupta, MD Dickey, MJ Butte and GM Whitesides. Thread as a matrix for biomedical assays. ACS Appl. Mater. Interfac. 2010; 2, 1722-8.

A Nilghaz, DR Ballerini and W Shen. Exploration of microfluidic devices based on multi-filament threads and textiles: A review. Biomicrofluidics 2013; 7, 051501.

S Sateanchok, S Wangkarn, C Saenjum and K Grudpan. A cost-effective assay for antioxidant using simple cotton thread combining paper-based device with mobile phone detection. Talanta 2018; 177, 171-5.

P Jarujamrus, N Malahom, S Puchum, R Meelapsom, M Amatatongchai, A Siripinyanond, S Chairam and C Kulsing. Complexometric and argentometric titrations using thread-based analytical devices. Talanta 2018; 183, 228-36.

S Farajikhah, JM Cabot, PC Innis, B Paull and G Wallace. Life-saving threads: Advances in textile-based analytical devices. ACS Comb. Sci. 2019; 21, 229-40.

LS Clesceri, AE Greenberg and AD Eaton. 4500-P E. Ascorbic acid method. In: APHA-AWWA-WEF (Ed.). Standard methods for the examination of water and wastewater analysis. 23th ed. American Public Health Association, Washington DC, 2017, p. 524-5.

A Gonzalez, M Gaines and FA Gomez. Thread‐based microfluidic chips as a platform to assess acetylcholinesterase activity. Electrophoresis 2017; 38, 996-1001.

A Walters, D Santillo and P Johnston. An overview of textiles processing and related environmental concerns. University of Exeter, Exeter, England, 2005.

JN Miller and JC Miller. Statistics and chemometrics for analytical chemistry. Prentice Hall, New Jersey, 2010.

Downloads

Published

2022-08-13

How to Cite

Sereenonchai, K. ., Inla, C. ., & Buajarern, S. . (2022). Thread-Based Analytical Device for Determination of Phosphorus Content in Fertilizers. Trends in Sciences, 19(16), 5616. https://doi.org/10.48048/tis.2022.5616